Page 48
•
Heating:
The chamber is warmed up until the Keep-Heat
temperature is reached inside the chamber.
•
Keep Heat:
Keep-Heat temperature is maintained constant for the
preset Keep-Heat time.
•
Cooling:
N/A.
•
Fast Exhaust:
The steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
•
Drying:
N/A
•
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
Note:
Some water may remain in the chamber.
5.13
Program 13: Air Steam Mixture
This program is intended for liquids in soft packages when the
manufacturer recommends autoclaving at temperature of 121ºC
(250
F) for 15 minutes.
Sterilization is controlled in a way that the sterilization pressure
remains approx. 30% above the theoretical pressure corresponding to
the same temperature according to the steam table. These intends for
prevent swelling or warping of the package.
Cautions!
Both PT100 temperature sensors must be inside the bottles.
For proper sterilization, fill the bottles with approximately the
same amount of liquid.
Nominal parameters default settings
•
Sterilization temperature: 121
C (250
F).
Содержание LabSci 11-V
Страница 19: ...Page 18 8 Right service door lock...
Страница 22: ...Page 21 10 Drain outlet...
Страница 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Страница 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Страница 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Страница 79: ...Page 78...
Страница 90: ...Page 89 Attention Use protective gloves in order not to burn your hands with the hot steam Safety valve ring...