Page 94
Message /
Symbol /
Problem
Failure Description
Corrective Action
sensor or Pressure sensor is
disconnected or out of range.
Chamber
temperature
not in range
This message is displayed if
the temperature in the
chamber is too high or too
low from the normal range.
Call for service.
Chamber
pressure not
in range
This message is displayed if
the pressure in the chamber
is too high or too low from the
normal range.
Call for service.
I/O Card
Failed
This message is displayed if
I/O card is faulty (both while
cycle is running or not).
Call for service.
I/O card is
not
connected
This message is displayed if
I/O card is disconnected
(both while cycle is running
or not).
Call for service.
Low
Vacuum’ is
displayed
Message is displayed and
FAIL indicator lights if in the
air removal stage, a vacuum
level of 15kPa is not reached
during 20minutes after the
cycle is started
1
Perform a new
cycle.
2
Call the
technician.
3
The bio-hazard
filter may be clogged.
Since the door cannot
be opened until a
complete successful
cycle is completed
– call
for technical service.
Содержание LabSci 11-V
Страница 19: ...Page 18 8 Right service door lock...
Страница 22: ...Page 21 10 Drain outlet...
Страница 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Страница 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Страница 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Страница 79: ...Page 78...
Страница 90: ...Page 89 Attention Use protective gloves in order not to burn your hands with the hot steam Safety valve ring...