Page 91
Message /
Symbol /
Problem
Failure Description
Corrective Action
damaged.
Low Temp
This message is displayed if
the temperature drops for
more than 1 second below
the sterilization temperature
during sterilization cycle.
Perform a new cycle.
High Temp
This message is displayed if
the temperature raises 7°F
(4°C) above sterilization
temperature during the
sterilization stage for 2
seconds during sterilization
cycle.
Perform a new cycle.
High Temp.
(Ending)
This message is displayed if
the system cannot reach the
required temperature, in the
chamber, within 10 minutes.
Perform a new cycle.
High Temp.
(Cooling)
This message is displayed if
the system cannot reach the
required temperature, in the
cooling stage, within preset
time.
Check and fix the city
(tap) water supply.
Heat Time
Error
This message is displayed if
the system cannot reach the
required temperature, in the
chamber, within the preset
time.
Verify that the autoclave
is not overloaded.
Heat Time
Error (Keep)
This message is displayed if
the system cannot reach the
required temperature, in the
chamber, during the optional
"Keep Heat" stage, within the
preset time.
Verify that the autoclave
is not overloaded.
Low
Pressure
This message is displayed if
Chamber Pressure drops
below the sterilization
pressure (134°C = 304 kPa,
121°C = 205 kPa) for 2
seconds during the
sterilization stage.
Perform a new cycle.
Low
Pressure
(Cooling)
This message is displayed if
the pressure in chamber does
not reach the preset pressure
before initiating the cooling
Check and fix the
compressed air supply.
Содержание LabSci 11-V
Страница 19: ...Page 18 8 Right service door lock...
Страница 22: ...Page 21 10 Drain outlet...
Страница 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Страница 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Страница 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Страница 79: ...Page 78...
Страница 90: ...Page 89 Attention Use protective gloves in order not to burn your hands with the hot steam Safety valve ring...