Page 9
6.
Individual glass bottles may be placed within an appropriate container
that will be placed in a basket. Never place glass bottles on the floor
of the autoclave. Never fill more than 2/3 of the bottle volume.
7.
On closing the autoclave's door, make sure it is properly locked
before activating.
8.
Before withdrawing baskets, wear heat resistant gloves.
9.
Before opening the door, verify that there is no pressure in the
chamber (chamber pressure gauge is located on the autoclave's front
panel or door, depends on model).
10.
Open the door slowly to allow steam to escape and wait 5 minutes
before you remove the load. When sterilizing liquids, wait 10 minutes.
11.
Once a month, ensure that the safety valves are functioning, and
once annually a certified tester must conduct pressure chamber
safety tests.
12.
Once annually, or more frequently, effective tests must be performed,
i.e., calibration and validation.
13.
Examine the condition of assemblies on a regular basis. Make sure
there are no leaks, breaks, blockages, whistles, or strange noises.
14.
It is required to conduct maintenance operations as instructed.
15.
Immediately notify the person in charge of any deviation or risk for the
proper function of the device.
16.
For WR models, drain the water from the mineral free water reservoir
once a week, and refill with fresh mineral-free water or distilled water
(see 10.1).
Содержание LabSci 11-V
Страница 19: ...Page 18 8 Right service door lock...
Страница 22: ...Page 21 10 Drain outlet...
Страница 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Страница 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Страница 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Страница 79: ...Page 78...
Страница 90: ...Page 89 Attention Use protective gloves in order not to burn your hands with the hot steam Safety valve ring...