Page 43
•
Sterilization time: 45 minutes
•
Drying time: 1 minute. (PV
– pre-vacuum models only).
Operations sequence:
•
(PV, pre-vacuum models only)
Pulse low/Pulse high:
At one
pulse it will build vacuum down to 25 kPa.
All exhaust from the
chamber is performed through the bio-hazard filter.
•
Water Inlet:
Water enters the chamber
•
Heating:
The chamber is heated by actuation of electrical heaters
until the sterilization temperature is reached.
•
Sterilization:
Sterilization temperature is maintained constant during
the sterilization time.
•
Cooling:
N/A
•
Fast Exhaust:
The steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
•
Note:
If the cycle fails, fast exhaust is performed through the bio-
hazard filter. Some water may remain in the chamber after a failed
cycle. To remove such water, press Start and repeat the cycle. If the
problem persists, call for service.
•
(PV, pre-vacuum models only)
Drying
: Vacuum is built up and the
air pump is working.
•
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.9
Program 9: Bio Hazard Liquids (BH, bio-hazard
models only)
All exhaust from the chamber before completion of the
Sterilization stage is performed through the bio-hazard filter
. For
liquids when the manufacturer recommends autoclaving at
temperatures of 121ºC (250
F) for 45 minutes.
Содержание LabSci 11-V
Страница 19: ...Page 18 8 Right service door lock...
Страница 22: ...Page 21 10 Drain outlet...
Страница 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Страница 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Страница 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Страница 79: ...Page 78...
Страница 90: ...Page 89 Attention Use protective gloves in order not to burn your hands with the hot steam Safety valve ring...