Page 50
•
Sterilization temperature: 121
C (250
F).
•
Sterilization time:
20
minutes.
Operations sequence:
•
Heating:
Steam enters the chamber and heats it up until the
sterilization temperature is reached.
During the interval of the heating stage, between 100°C and
121°C, the heating rate is kept to approximately a rise of 1 degree
per minute. This will make the last stage of the heating take
approximately 21 minutes. This is to check the durability of the
bottles.
•
Sterilization:
Sterilization temperature is maintained constant during
the sterilization time.
•
Cooling:
Forced cooling to the required end temperature, chamber
pressure remains at approx. 300 kPa and the cooling valve is opened.
During the interval of the cooling stage, between 121°C and
100°C the cooling rate is kept to approximately a decrease of 1
degree per minute. This will make the last stage of the cooling
take approximately 21 minutes. This is to check the durability of
the bottles.
•
Fast Exhaust:
The steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
•
Drying:
N/A
•
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.15
Program 15: Durham
For liquids when the manufacturer recommends autoclaving at
temperatures of 121ºC (250
F) for 15 minutes.
Содержание LabSci 11-V
Страница 19: ...Page 18 8 Right service door lock...
Страница 22: ...Page 21 10 Drain outlet...
Страница 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Страница 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Страница 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Страница 79: ...Page 78...
Страница 90: ...Page 89 Attention Use protective gloves in order not to burn your hands with the hot steam Safety valve ring...