Page 88
Use protective gloves in order not to burn your hands with the hot
steam.
11.2.2
ASME-approved type safety valve
Operate the sterilization cycle according to the manual.
Allow a pressure of approximately 200 kPa (29-psi) to build up in the
chamber.
Operate the safety valve by pulling the ring of the safety valve using a
tool, i.e. screwdriver, hook etc. and lift the safety valve ring for 2
seconds. Be careful not to burn your hands.
Press the
STOP/STOP
key to interrupt the operation, and exhaust
steam from chamber.
Wait until pressure goes down to zero, only then can the door be
opened.