Page 63
Below is t
he instruction for autoclave’s maintenance menu.
6.7.1
Reset atmospheric pressure
In this menu you can reset the atmospheric pressure value. To do so:
1.
Browse to the following folder:
Maintenance\Reset atmospheric pressure
The following screen will appear:
2.
Leave the door open for 2 minutes at least. Ambient temperature
should be less than 45°C.
Note:
Please reset the atmospheric pressure when you install the autoclave
for the first time, and each time you relocate or calibrate the
autoclave.
6.7.2
Printer test
In this menu you can check the normal function of the printer. The
printer will print the list of errors.
Browse to the following folder:
Maintenance\Printer test