Page 93
Message /
Symbol /
Problem
Failure Description
Corrective Action
Air Error
This message is displayed at
the end of the cycle If the
autoclave does not reach the
atmospheric pressure after
10 minutes.
Wait until the autoclave
reaches the atmospheric
pressure and perform a
new cycle.
Compressed
air supply
error
This message is displayed in
case of a compressed air
supply malfunction.
Check and fix the
compressed air supply.
Periodical
check time
exceeded -
Please call
for service
The periodical maintenance
time has passed.
Call for service.
Cycle
counter
exceeded -
Please call
for service
Number of cycles, since last
periodical maintenance,
exceeded the preset number
as defined by "cycle counter"
parameter.
Call for service.
Power Down
This message is displayed if
power down has occurred
during the cycle. (this
message will print out in the
printer after the autoclave will
turn on).
Turn on the autoclave
and wait until the
autoclave is ready
(reaches the safe
condition) and perform a
new cycle.
Supply
distilled
water error
(digitat input
option)
This message is displayed in
case of a mineral free water
supply malfunction.
Check and fix the
mineral free water
supply
Supply water
error (digitat
input option)
This message is displayed in
case of a city (tap) water
supply malfunction.
Check and fix the city
(tap) water supply.
Compressed
air supply
error (digitat
input option)
This message is displayed in
case of a compressed air
supply malfunction.
Check and fix the air
supply.
No Water
This message is displayed if
the electrode in the chamber
did not sense water within
the preset time.
1.
check and fix the
mineral free water
supply.
2.check and clean the
water inlet filter.
3.
Clean the water
level electrode.
Analog Input
Error
This message is displayed
when any Temperature
Call for service.
Содержание LabSci 11-V
Страница 19: ...Page 18 8 Right service door lock...
Страница 22: ...Page 21 10 Drain outlet...
Страница 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Страница 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Страница 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Страница 79: ...Page 78...
Страница 90: ...Page 89 Attention Use protective gloves in order not to burn your hands with the hot steam Safety valve ring...