TD-700 Laboratory Fluorometer Operating Manual
Page 5
I.
Introduction
A.
Description
The TD-700 Fluorometer is a compact, laboratory fluorometer designed for
discrete sample measurement of various fluorescent materials, including
chlorophyll, rhodamine, fluorescein, histamine, vitamins, and other fluores-
cent compounds. When properly calibrated, TD-700 will electronically set
the optimal sensitivity and range for your samples, and will readout the
actual concentration of the compound when in the Direct Concentration
mode. The TD-700 can be set up to detect the specific compound you
want to measure simply by changing the lamp and optical filters.
B.
Inspection and Setup
Upon receiving your fluorometer, please inspect it carefully and make sure
all accessories are present (refer to the packing list shipped with the
instrument).
To get started (Refer to the Quick View Diagram, page 7):
1.
Locate the instrument on a flat surface with the back of the instrument
well ventilated (at least three inches from a wall).
2.
Open the sample chamber lid located on the top of the unit.
3.
Remove the Filter Cylinder by grasping the inside rim of the cylinder
and pulling up.
4.
Install the correct optical filters for your application. See Section III for
details.
5.
Check that the correct lamp is installed for your primary application.
See Section IV for details.
6.
Check to see that the sample adaptor is the correct one for your
needs. Refer to Appendix 6 for a detailed explanation of TD-700
Sample Adaptors.
7.
Reinsert the Filter Cylinder, lining up either the A, B, C or D applica-
tion indicators on the cylinder with the silver alignment mark on the
inside rim of the sample chamber. Replace the sample adaptor making
sure that the adaptor is properly seated in the filter cylinder, and close
the sample chamber lid.
8.
Plug in the unit.
Be sure to use only the Turner Designs power
supply provided with your instrument
. It is critical to use the power
supply provided by Turner Designs to meet EMI requirements.
9.
Turn on the power switch (rear of unit) and allow the unit to warm up for
the countdown period (600 seconds; 10 minutes).
NOTE: For optimal instrument stability, leave the instrument turned on.
TD-700 Front View
TD-700 Side View
Filter Cylinder
Assembly
TD-700 Top View
Содержание TD-700
Страница 5: ...TD 700 Laboratory Fluorometer Operating Manual Page 4 NOTES ...
Страница 16: ...TD 700 Laboratory Fluorometer Operating Manual Page 15 NOTES ...
Страница 22: ...TD 700 Laboratory Fluorometer Operating Manual Page 21 NOTES ...
Страница 28: ...TD 700 Laboratory Fluorometer Operating Manual Page 27 NOTES ...