TD-700 Laboratory Fluorometer Operating Manual
Page 45
APPENDIX 8
Error Messages and Notes
During calibration, the TD-700 will display an error message if incorrect values
are entered or if the readings may result in an inadequate calibration.
A.
Invalid Input
This screen will appear when you enter a value that is outside the range for
the parameter. For example, if you enter a number greater than 1000 for
the Max range, you will see “Invalid Input.”
B.
Error1
The concentration entered for the standard (Std) is greater than the
maximum concentration entered for the Maximum Range. Press <ENT>
and enter a lower concentration for the Std. OR press <ESC>, then <1> to
abort the calibration; use a lower concentration standard or enter a higher
value for the Maximum Range.
C.
Note 1
If the sensitivity exceeds the maximum before reaching the target sensitiv-
ity, a note message will appear. The unit will automatically adjust sensitiv-
ity to the maximum. If sensitivity falls below the minimum before reaching
the target sensitivity, the unit will automatically adjust sensitivity to the
minimum. Note that this may change the Max range concentration you
entered.
D.
Error3
The
concentration entered
for #2-5 Std is greater than the concentration
entered for the highest Std. Press <ENT> and enter a lower concentration
for the Std. OR press <ESC>, then <1> to abort the calibration. Begin
again using the highest concentration standard first.
E.
Error4
Fluorescence reading
for #2-5 Std is higher than the reading for the
highest standard. Press <ESC> to abort the calibration; rerun using the
highest concentration standard first.
F.
Error5
Fluorescence reading
for blank is higher than the reading for one or more
of the standards. Press <ENT> to rerun blank; make sure blank is OK and
rerun. OR, press <ESC> to abort the calibration; calibrate again using
standards more concentrated than blank.
Invalid input
<ENT> to reenter
A.
Error1: Std conc
> max range
B.
Note1: MAX (or MIN) sens
Press <ENT>
C.
D.
E.
Error3: Std conc
ent. >Hi Std
Error4: #X Std
reads >Hi Std
Error5: Blk reads
> Std #X
F.
Содержание TD-700
Страница 5: ...TD 700 Laboratory Fluorometer Operating Manual Page 4 NOTES ...
Страница 16: ...TD 700 Laboratory Fluorometer Operating Manual Page 15 NOTES ...
Страница 22: ...TD 700 Laboratory Fluorometer Operating Manual Page 21 NOTES ...
Страница 28: ...TD 700 Laboratory Fluorometer Operating Manual Page 27 NOTES ...