TD-700 Laboratory Fluorometer Operating Manual
Page 34
C.
Calibration Printout Using the Direct Concentration
Calibration Procedure:
(Example only! Your numbers will vary.)
Press <D> from the Setup/Cal screen to send the calibration information
to the printer or computer.
Mode: Multi-Optional
Cal. Procedure: Direct Conc.
Units: ppb
Sensitivity Level: 48
Current Calibration
Entered Max Range: 220.00
Calibration Standard Summary:
Std.
Conc.
Raw Fluor.
BLK
0
0
Std.3
1.50
6.706
Std.2
6.20
29.09
Hi Std.
50.00
227.9
Calibration Curve Data:
Conc. Range
Fluor. Range
Slope
0(BLK)
-
1.500
0.00
-
6.710
0.223
1.500
-
6.200
6.710
-
29.09
0.210
6.200
-
50.00
29.09
-
227.9
0.220
Slope %RSD: 3.127
An Explanation of the Printout
The first four lines of the printout display the TD-700's current parameters:
the mode, the cal procedure, the units, and the instrument sensitivity level.
The sensitivity level (a number between 1 and 100) is a measure of the
photomultiplier voltage. Next is the calibration that is stored in the unit’s
memory. The maximum concentration range entered during the calibration
sequence is displayed here. Following this is the Calibration Standard
Summary which displays each standard, its concentration, and its corre-
sponding raw fluorescence value. Concentration and raw fluorescence
values are organized in the Calibration Curve Data table so that the slope
of the line between each data point is displayed. The percent relative
standard deviation (%RSD) is displayed below this. The % RSD is an
indicator of how well each of the slope values correlate to one another. A
consistent slope (a slope that is almost the same for all data points) will
result in a low %RSD. A nonlinear set of data will result in a high %RSD.
The %RSD can be thought of as an overall indicator of the data quality. A
%RSD of less than 25% is considered by many to be acceptable while a
%RSD of less than 5 indicates excellent linearity. Both the individual
slopes and average slope will change depending on the numbers input
during calibration. There is no "best" slope value.
Содержание TD-700
Страница 5: ...TD 700 Laboratory Fluorometer Operating Manual Page 4 NOTES ...
Страница 16: ...TD 700 Laboratory Fluorometer Operating Manual Page 15 NOTES ...
Страница 22: ...TD 700 Laboratory Fluorometer Operating Manual Page 21 NOTES ...
Страница 28: ...TD 700 Laboratory Fluorometer Operating Manual Page 27 NOTES ...