TD-700 Laboratory Fluorometer Operating Manual
Page 20
1. Setup
2. Calibration
1. Setup
2. Calibration
Sample = XXX
<ENT>
Sensitivity Set
Sens Factor: XX
SETTING Sens
Sens Factor: XX
Insert Typical
Sample <ENT>
Blank: X.X
Cal Std: XXX.X
Calibration: Simple Mode
1. Mode
<Simple>
Multi-Optional
2.
4.
7.
5.
1.
<1> - Abort Cal
<ESC> - Resume
9.
VII. Calibration: Simple Mode
The Simple Mode calibration procedure is a single point calibration in which a
standard is run in order to set the optimal range and sensitivity of the instrument.
The chosen sample will be set automatically to one half of the maximum value that
can be accurately read by the instrument. You cannot manually adjust the
instrument range and sensitivity during this procedure. Also, there is no option to
read and subtract a blank. If you do not require these capabilities, you may wish
to use the Simple Mode calibration as it is expedient and straightforward.
1.
To choose the Simple Mode, press <ENT> from the HOME screen, press
<1> for Setup, then <1> again for Mode. Use <
↔
>
to choose the Simple
Mode. Press <ESC> twice to return to the Setup/Cal Screen.
2.
Press <2> from the Setup/Cal Screen. The Simple Mode calibration
sequence will appear.
3.
Fill a clean test tube or cuvette with a sample that is about half the
maximum concentration you wish to read. You do not need to know the
exact concentration; you are using it to set the optimal instrument range
and sensitivity. Wipe the outside of the test tube or cuvette dry, and insert
it into the sample adaptor in the sample chamber.
4.
The TD-700 will now set sensitivity, as indicated by the SENS FACTOR
(sensitivity factor), so that the sample you inserted will read 500 (half of
maximum) on the HOME screen.
5.
Upon finishing the sensitivity adjustment, the screen prompts you to
acknowledge the value set by pressing <ENT>. Then, calibration data will
automatically printout to a printer or a computer and return you to the HOME
screen. See Appendix 2 for details regarding the calibration printout.
6.
Errors: After reading the sample, the TD-700 will automatically adjust to the
optimal range for sample measurement. If the sample used is too
concentrated or too dilute, the instrument may not be able to reach the target
sensitivity. In these cases, a note message will appear indicating that, based
on the calibration, the unit has reached its sensitivity maximum or minimum.
By pressing <ENT> you tell the instrument to accept the maximum or
minimum sensitivity value. It is recommended, however, that you adjust your
sample concentrations to fall within range, and then recalibrate.
7.
To view the last calibration data set, press <ENT> from the HOME screen,
then <9> to view your blank and calibration standard value. Press <H> to
return to the HOME screen.
8.
To print the last calibration data set, press <ENT> from the HOME screen,
then <D>. Press <H> to return to the HOME screen.
9.
To abort the calibration, press <ESC> at any time during the calibration
sequence. Press <1> to abort or <ESC> to resume.
Содержание TD-700
Страница 5: ...TD 700 Laboratory Fluorometer Operating Manual Page 4 NOTES ...
Страница 16: ...TD 700 Laboratory Fluorometer Operating Manual Page 15 NOTES ...
Страница 22: ...TD 700 Laboratory Fluorometer Operating Manual Page 21 NOTES ...
Страница 28: ...TD 700 Laboratory Fluorometer Operating Manual Page 27 NOTES ...