TD-700 Laboratory Fluorometer Operating Manual
Page 35
Note that during calibration, the maximum you can enter for the Max
Range is 1000. However, the actual maximum achieved after calibration
may be higher, depending upon the reading for the Blank and for the High
Standard. This is a normal condition; the instrument will set to the best
sensitivity to obtain the Max Range you specified.
D.
Calibration Printout in the Multi-Optional Raw
Fluorescence Mode:
(Example only! Your numbers will vary.)
Press <D> from the Setup/Cal screen to send the calibration information
to the printer or computer.
Mode: Multi-Optional
Cal. Procedure: Raw Fluor.
Units: fsu**
Sensitivity Level: 48
RAW Fluor. Blank = 500.0
RAW Fluor. Blk = 123.0
Adj. Fluor. Sample = 377.0
Adj. Fluor. Blank = 0.000
An Explanation of the Printout
The first four lines of the printout display the TD-700's current parameters:
the mode, the cal procedure, the units, and the instrument sensitivity level.
Following this is the calibration stored in the unit’s memory. The calibration
stored consists of the raw and adjusted sample and blank value (if appli-
cable), expressed in raw fluorescence units.
E.
Calibration Printout Using the Simple Mode:
(Example only!
Your numbers will vary.)
Press <D> from the Setup/Cal screen to send the calibration information
to the printer or computer.
Mode: Simple
Cal. Procedure: Raw Fluor.
Units: fsu**
Sensitivity Level: 48
RAW Fluor. Sample = 500.0
An Explanation of the Printout
The first four lines of the printout display the TD-700's current parameters:
the mode, the cal procedure, the units, and the instrument sensitivity level.
Following this is the calibration stored in the unit’s memory. The calibration
stored consists of the sample value only, expressed in raw fluorescence
units.
** fsu = Fluorescent Signal Units, the
designated units for Raw Fluorescence
readings.
Содержание TD-700
Страница 5: ...TD 700 Laboratory Fluorometer Operating Manual Page 4 NOTES ...
Страница 16: ...TD 700 Laboratory Fluorometer Operating Manual Page 15 NOTES ...
Страница 22: ...TD 700 Laboratory Fluorometer Operating Manual Page 21 NOTES ...
Страница 28: ...TD 700 Laboratory Fluorometer Operating Manual Page 27 NOTES ...