TD-700 Laboratory Fluorometer Operating Manual
Page 13
C. Setup/Calibration Screen
1. Setup
2. Calibration
XXX ppm
<ENT> -Setup & Cal
B. HOME Screen
TURNER DESIGNS
T7-1B 10/96 600
A. Power-up Screen
V.
Instrument Parameters (Firmware)
Instrument parameters are set through the TD-700 firmware interface.
Firmware screens are called up using the keypad. To see how thescreens
fit together, see the firmware flowchart on pages 16 & 17. For details about
alarms and instrument diagnostics, see Appendix 3, Section A, and
Section B.
A.
Power-up Screen
When the unit is first turned on, a screen appears showing the title,
software version, and a 600-second (10-minute) countdown. Once the 600-
second countdown expires, the HOME screen will appear. The countdown
allows the instrument to warm up adequately before measurements begin.
For optimal stability and accuracy, it is recommended that you allow the
instrument to complete the countdown period. For best results, leave the
instrument turned on in your laboratory.
If you wish to bypass the countdown period, press <H>, <ESC>, or
<ENT>. From the HOME screen, press <9> to return to the Power-up
screen.
B.
HOME Screen
After the countdown period, the HOME screen is displayed. The HOME
screen is where samples are read and data is sent to the printer or
computer. Press <H> to return to this screen from any screen (except
during calibration).
From the HOME screen, several functions can be accessed.
- Press <D> to send data to a printer or computer.
- Press <*> to Discrete Sample Average (See Section X for details).
- Press <0> to Autozero or blank the instrument (See Section X for
details).
- Press <8> for Diagnostic Screens.
- Press <7> for Data Stream.
C.
Setup/Calibration Screen
The Setup/Calibration screen, accessed from the HOME screen by
pressing <ENT>, is a screen where the user selects either to go to the
calibration sequence or to review or change the setup parameters. The
user may also press <D> from this screen to send the current calibration
information to a printer or computer.
Содержание TD-700
Страница 5: ...TD 700 Laboratory Fluorometer Operating Manual Page 4 NOTES ...
Страница 16: ...TD 700 Laboratory Fluorometer Operating Manual Page 15 NOTES ...
Страница 22: ...TD 700 Laboratory Fluorometer Operating Manual Page 21 NOTES ...