TD-700 Laboratory Fluorometer Operating Manual
Page 12
TD-700 with Halogen Lamp
Rear View
Quick-release
latches
B.
The Quartz Halogen Lamp
The average halogen lamp life is 2000 hours.
Check the lamp if:
1.
The instrument is not responding, even though the unit is plugged in
and the power is on
2.
The readings are very low, unstable, or drifting.
To replace the quartz halogen lamp:
1.
Turn off the instrument and remove the lamp access panel by pulling
on the two black quick-release latches on the rear panel of the instru-
ment.
2.
Unplug the power jack that connects the lamp panel assembly to the
instrument. The lamp access panel should now pull free of the instru-
ment.
3.
DO NOT TOUCH THE LAMP BULB with an ungloved hand; the
bulb may be hot, and/or the oils from your hand will affect the
lamp's transmittance and cause incorrect readings
. If the bulb is
burned out, grab the bulb with a tissue and pull it out of the assembly.
Using the same tissue, insert a new bulb (orientation does not matter).
4.
Plug the power jack back into the main instrument and replace the
lamp access panel. Push in the black quick-release latches to snap
the panel in place.
5.
Turn on the power. Check to ensure the lamp is lit by viewing it
through the fan opening. Also make sure the fan is operating. If the fan
is not operating, check the connector plug.
6.
You must recalibrate if you replace the lamp.
Содержание TD-700
Страница 5: ...TD 700 Laboratory Fluorometer Operating Manual Page 4 NOTES ...
Страница 16: ...TD 700 Laboratory Fluorometer Operating Manual Page 15 NOTES ...
Страница 22: ...TD 700 Laboratory Fluorometer Operating Manual Page 21 NOTES ...
Страница 28: ...TD 700 Laboratory Fluorometer Operating Manual Page 27 NOTES ...