TD-700 Laboratory Fluorometer Operating Manual
Page 38
APPENDIX 4
Troubleshooting
Symptom
Solution
Section To See
Power ON but
Check the lamp; change if necessary.
Section IV
readings stay
Make sure the sample lid is closed
at zero
tightly.
Power switch
Make sure unit is plugged in securely;
ON, but no
check power from the wall socket; Make
power
sure power supply is functioning.
Display reads
Sample reading is too high for the
Appendix 8, H.
OVER
instrument to read at the current
sensitivity level. Dilute samples to get
an on-scale reading. Or recalibrate and
reduce sensitivity.
Display reads
Reading for Blank is less than zero
Appendix 8, H.
negative (-)
due to drift; reading for samples are
less than Blank. Reset Blank to Zero
or recalibrate.
Drifting
The instrument is open loop and
Section VII
readings
readings will drift over a period of time;
Section VIII
recalibrate to cure.
Section IX
Erratic readings
Instrument readings will fluctuate
Section III
somewhat due to instrument noise,
Section IV
especially at high sensitivity and when
reading low concentrations. If readings
are very erratic, check the lamp and
optical filters for damage/freckling.
Low readings;
Sensitivity was not adjusted adequately
Section VII
not enough
during calibration; recalibrate. Or, make
Section VIII
resolution
sure the proper lamp and filters are
Section IX
being used.
Section III
Section IV
Содержание TD-700
Страница 5: ...TD 700 Laboratory Fluorometer Operating Manual Page 4 NOTES ...
Страница 16: ...TD 700 Laboratory Fluorometer Operating Manual Page 15 NOTES ...
Страница 22: ...TD 700 Laboratory Fluorometer Operating Manual Page 21 NOTES ...
Страница 28: ...TD 700 Laboratory Fluorometer Operating Manual Page 27 NOTES ...