TD-700 Laboratory Fluorometer Operating Manual
Page 22
VIII. Calibration: Multi-Optional Mode - Raw
Fluorescence
The Multi-Optional Mode - Raw Fluorescence calibration procedure is a single
point calibration in which a standard and optional blank are run in order to set the
optimal range and sensitivity of the instrument. The user chooses from a range of
100-950 on a scale of 1000, the relative magnitude of the standard. The user
also has the option to subtract the blank signal from all the readings.
1.
To choose the Multi-Optional - Raw Fluorescence Mode, press <ENT>
from the HOME Screen, press <1> for Setup, then <1> again for Mode.
Use <
↔
>
to choose the Multi-Optional Mode. Press <ESC> to return to
the previous screen, then press <2> to choose the calibration procedure.
Use the <
↔
>
key to choose “Raw Fluor.” for the Raw Fluorescence
calibration procedure. Press <ESC> twice to return to the Setup/Cal
Screen.
2.
To access the calibration sequence, press <2> from the Setup/Cal
Screen. The Multi-Optional - Raw Fluorescence calibration sequence will
appear.
3.
Fill a clean test tube or cuvette with a sample that you wish to read. You
do not need to know the exact concentration; you are using it to set the
optimal instrument sensitivity and range. Wipe the outside of the test tube
or cuvette dry, and insert it into the sample adaptor in the sample
chamber. Press <ENT> to proceed to the next screen.
4.
Default setting raw fluorescence reading is 800. If this is acceptable press
<1> to continue. If you want to change this value, press <9>. Type in the
new value (between 100 - 950) and press <ENT>. Press <1> to continue.
5.
The TD-700 will now set its sensitivity, as indicated by the SENS FACTOR
(sensitivity factor), based on the final sample value that you accepted.
Once the sample is set, the TD-700 asks whether you want to run a blank.
If you wish to have blank subtracted, press <1>. If not, press <9> and the
calibration sequence will end here and you will be returned to the HOME
screen. At this point, the calibration data will printout automatically to a
printer or a computer. See Appendix 2 for details regarding the calibration
printout.
Calibration: Multi-Optional
Raw Fluorescence
1. Setup
2. Calibration
1. Mode
2. Cal Procedure
Simple
<Multi-Optional>
1. Mode
2. Cal Procedure
<Raw Fluor.>
Direct Conc.
1. Mode
2. Cal Procedure
1. Setup
2. Calibration
Insert Typical
Sample <ENT>
SETTING Sens
Sens Factor: XX
Sensitivity Set
Sens Factor: XX
Sample = XXX
<ENT>
Read & Subtract
Blank? 1=Yes 9=No
1.
2.
3.
4.
Содержание TD-700
Страница 5: ...TD 700 Laboratory Fluorometer Operating Manual Page 4 NOTES ...
Страница 16: ...TD 700 Laboratory Fluorometer Operating Manual Page 15 NOTES ...
Страница 22: ...TD 700 Laboratory Fluorometer Operating Manual Page 21 NOTES ...
Страница 28: ...TD 700 Laboratory Fluorometer Operating Manual Page 27 NOTES ...