TD-700 Laboratory Fluorometer Operating Manual
Page 29
B.
Resetting Blank to Zero
This feature allows you to reset the blank to zero from the HOME screen
when in the Multi-Optional Mode. Although running a calibration is the best
method to minimize sample error due to instrument drift, a quick and easy
way to reduce the effects of dark current, and blank fluorescence, is to
"autozero" or "reset the blank to zero" from the HOME screen. However,
be sure to run at least one calibration before using this feature. Resetting
the blank to zero from the HOME screen is quite useful if you wish to read
multiple sets of blank and relative sample readings, and do not wish to
recalibrate. When resetting the blank to zero, you are replacing the last
blank setting with the new blank setting. Again, this procedure will not
compensate for errors in sample readings due to instrument drift.
For
maximum accuracy, you should recalibrate.
1.
Insert a blank, then press <0> from the HOME screen.
2.
Press <9> to abort the procedure or <1> to Reset Blank to Zero.
Press <1> and without further prompting, the instrument will blank to
zero.
3.
If the reading for the blank exceeds the maximum, the unit will display
an error message: “Blank level too high.” If this message should
appear, evaluate your sample to ensure that you are reading the blank
and not another type of sample. Once you have verified that you are
reading the blank, try the procedure again. If the message persists,
recalibrate the instrument.
C.
Data Stream
This feature allows you to output data directly to a printer or a computer at
regular intervals. Only the value seen on the LCD screen will be printed. No
index or raw fluorescence value will be printed.
1.
Press <7> from the HOME screen.
2.
Press <
↔
> to toggle between <Yes> and <No> (turns data stream on
or off).
When the data stream feature is on, the data rate can be changed. The
number selected is the time interval between data output.
1.
Press <D> from the HOME screen when the data stream feature is on.
2.
Use the <
↔
> key to toggle between numbers.
BLK XXX ppm
<ENT> -Setup & Cal
Blank?
1. Yes
9. No
Data Stream?
Yes
<No>
Data Rate (sec)
<1>
2
5
10
30
60
Содержание TD-700
Страница 5: ...TD 700 Laboratory Fluorometer Operating Manual Page 4 NOTES ...
Страница 16: ...TD 700 Laboratory Fluorometer Operating Manual Page 15 NOTES ...
Страница 22: ...TD 700 Laboratory Fluorometer Operating Manual Page 21 NOTES ...
Страница 28: ...TD 700 Laboratory Fluorometer Operating Manual Page 27 NOTES ...