TD-700 Laboratory Fluorometer Operating Manual
Page 14
D.
Setup Screen
This screen is used to choose the calibration procedure for the TD-700.
-Simple, Multi-Optional Raw Fluorescence, Multi-Optional Direct
Concentration. This screen will have 1 of 3 looks depending upon how the
instrument is currently configured. (See figure to the left).
To access the Setup Screen, from the Home screen press <ENT> then
<1>. Press the appropriate number to access the setup parameters.
1. Mode. This parameter allows the user to choose between Simple &
Multi-Optional calibration modes. (See Section V E for more details).
2. Cal Procedure. This parameter allows the user to choose between Raw
Fluorescence and Direct Concentration calibration procedures (See
Section V E for more details.
3. Units. Used only in Multi-Optional - Direct Concentration mode only.
This is to choose which units your standards and samples are
measured in.
For all 3 of these parameters, use the <-> key to toggle between choices.
When you have your choice selected, press <ESC> or <H>.
E.
Calibration Screens
The instrument will initiate the calibration process when <2> of the keypad
is pressed from the "Setup/Calibration" Screen. The firmware will guide
the user through the process. Depending upon the calibration mode and
procedure chosen in the Setup screens, the steps will be different. (See
firmware flowchart on pages 16 & 17). Here is a description of the 3
different calibration procedures.
Simple: One point calibration
This calibration allows you to use one standard (or sample) and no
blanks. The instrument gives the standard a relative value of 500 on a
scale of 0-1000. The instrument uses a preprogrammed absolute
zero as the 0 point.
Multi-Optional/ Raw fluorescence: two point calibration
(st blank)
This calibration allows you to use one standard and the option to blank
subtract. The user can specify the magnitude in relative fluorescence
units for the standard (on a scale of 0-1000). As an example, if you
expect your standard is ~75% of the full range for the application (see
application notes on our website for ranges), then call the standard
750 RFUs.
MultiOptional/Direct Concentration: Multi point calibration
(up to 5 sta blank)
This calibration allows you to use up to 5 different standards plus a
blank for the calibration. You can choose the units of your standards,
and define the magnitude of the maximum readable level. In the end,
the output can be in direct concentration so no further calculations are
needed.
1. Mode
2. Cal Procedure
Raw Fluorescence
1. Mode
Simple Mode
1. Mode 3. Units
2. Cal Procedure
Direct Concentration
Содержание TD-700
Страница 5: ...TD 700 Laboratory Fluorometer Operating Manual Page 4 NOTES ...
Страница 16: ...TD 700 Laboratory Fluorometer Operating Manual Page 15 NOTES ...
Страница 22: ...TD 700 Laboratory Fluorometer Operating Manual Page 21 NOTES ...
Страница 28: ...TD 700 Laboratory Fluorometer Operating Manual Page 27 NOTES ...