TD-700 Laboratory Fluorometer Operating Manual
Page 37
Raw
. The “raw” fluorescence signal output from the instrument’s light
detector (the photomultiplier tube). This is the output the TD-700 uses (in
conjunction with the actual value entered for the standards, and the
fluorescence signal for standards and blank) to arrive at the readout on the
HOME screen. It can be used to diagnose problems with the unit.
NOTE: To see the raw fluorescence for the standard(s) and the blank as
set during the last calibration, from the Calibration & Setup menu, press
<D> to print or send to a computer.
NOTE: If you calibrated in the Simple Mode or Multi-Optional Mode “Raw
Fluor.” Calibration Procedure, to see the raw fluorescence for the standard
and the blank as set during the last calibration, from the Calibration &
Setup menu, press <9> and they will appear on the display.
Operation Hour (Oper)
. Shows how many hours the instrument has been
operating.
Power level
. When at 100%, indicates power supply is functioning
correctly.