General Care, Troubleshooting and Service
6-31
Error
Code Error Message
Suggested Action
8
Sheath Flow Temperature Too Low
Check that flowmeter connector did not come loose during
shipping.
The flowmeter is reading outside its valid range of -10°C to
70°C. If the instrument is in a hot location, move it to a cooler
location. If not, the flowmeter may be defective.
The flowmeter is reading outside its valid range of -10°C to
70°C. If the instrument is in a cold location, move it to a
warmer location. If not, the flowmeter, its cable or connector
may be defective.
9
Sheath Flow Temperature Too High
10
Sample Flow Temperature Too Low
11
Sample Flow Temperature Too High
12
Charger Flow Temperature Too Low
13
Charger Flow Temperature Too High
14
Extraction Flow Temperature Too
Low
15
Extraction Flow Temperature Too
High
24
Positive Charger Current Too Low
Check the state of the Chargers
in the instrument’s User
Settings Window or view the Instrument Status window in the
EEPS software.
Check state of Flows in the instrument’s User Settings Window
or view the Instrument Status window in the EEPS software.
Chargers are turned off automatically when flows are turned
off.
The charger current is not stable or has hit a voltage limit to
meet the required current. Clean the charger assembly and
clean or replace the charger corona wire.
25
Positive Charger Current Too High
26
Positive Charger Voltage Too Low
27
Positive Charger Voltage Too High
28
Negative Charger Current Too Low
29
Negative Charger Current Too High
30
Negative Charger Voltage Too Low
31
Negative Charger Voltage Too High
34
Particle Concentration Too High
The concentration on one or more channels has exceeded its
upper limit. Lower the concentration of aerosol going into the
instrument. This is indicated in the software by a red box on
top of the column that is over-ranged in the 2D histogram.
34
Particle Concentration Too High
The RMS noise for a single channel is much larger than the
average of all the other channels. This could indicate that the
column is dirty. Clean the column or contact TSI.
50
Electrometer Channel High Offset
During instrument zeroing, the offset current for one or more
channels was higher than normal. This could indicate that the
column is dirty. View the last offset and RMS values to
determine the channel causing the error. If the channel is not
changing over time, the instrument may be okay to continue to
use. Otherwise, clean the column or contact TSI.
54
Electrometer Channel Noisy
During instrument zeroing, the RMS noise for one or more
channels was higher than normal. This could indicate that the
column is dirty. View the last offset and RMS values to
determine the channel causing the error. If the channel is not
changing over time, the instrument may be okay to continue to
use. Otherwise, clean the column or contact TSI.
Содержание SIZER EEPS 3090
Страница 2: ......
Страница 18: ...xvi Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 20: ...xviii Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 36: ...2 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 144: ...B 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 149: ...Serial Commands D 3 3 Enter the communications COM port 4 Enter the port settings described below and click OK...
Страница 184: ......