General Care, Troubleshooting and Service
6-5
4. The negative charger is located closest to the top of the charger column
(Figure 6-3). Remove the second set of screws (3) using an Allen
wrench (provided in the accessory kit). Loosen all screws ½ turn, then
remove each completely and put them aside for reassembly.
Figure 6-3
Remove the Screws of the Negative Charger Assembly
5. Carefully lift off the charger head (Figure 6-4).
Figure 6-4
Lift off the Negative Charger Head