6-12
Model 3090/3090AK Engine Exhaust Particle Sizer
TM
Spectrometer
5.
Unclip the wiring harness retainer using a flat-blade screwdriver, if
needed (Figure 6-11). The charger wiring harness should now be free.
Figure 6-11
Unclip the Wiring Harness Retainer
6.
Lift the charger head assembly off. It seals with an O-ring, so there will
be some resistance as you lift the head assembly.
7.
Turn the assembly over and set it off to one side, refer to
Figure 6-12.
8.
Locate the cleaning tool (TSI P/N 2900009
—assembly cleaning tool)
from the accessory kit and the 12 × 15 in. (30 × 38 cm) square
cleaning cloths. The cloths should be lint-free and may be dampened
with acetone if needed. Fold the cloth in the long direction to double up
the cloth (if this is too thick, cut the cloth in half to get a 7.5 × 12 in.
piece and try again).
9.
Apply a cleaning cloth to the cylinder as shown in Figure 6-12. One
edge should be attached to the loop fastener tape and the cloth
wrapped around the cylinder until it overlaps.
Figure 6-12
Cleaning Tool
Содержание SIZER EEPS 3090
Страница 2: ......
Страница 18: ...xvi Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 20: ...xviii Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 36: ...2 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 144: ...B 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 149: ...Serial Commands D 3 3 Enter the communications COM port 4 Enter the port settings described below and click OK...
Страница 184: ......