Software Operation
5-33
Table 5-3
Instrument Status Parameters
Parameter
Description
Sheath Flow
Volumetric flow rate of the particle free sheath air flow in
liter per minutes.
Sample Flow
Volumetric flow rate of the sample flow, in liter per
minutes. Sample flow plus extraction flow equals the
aerosol inlet flow.
Extraction Flow
Volumetric flow rate of the extraction flow in liter per
minutes.
Charger Flow
Volumetric flow rate of the particle free charger sheath
air in liter per minutes.
Internal Pressure
Absolute pressure in the sizing column in millibar. Used
for the volumetric flow calculation.
Column Voltage 1-3.
Voltage on the center electrode sections in Volt.
Column Voltage 1 = Top Section
Column Voltage 2 = Middle Section
Column Voltage 3 = Bottom Section
Sheath Temp.
Temperature of the particle free sheath air flow in
degree Celsius before it enters the sizing column.
Neg. Charger Current
and Voltage
Ion current of the negative charger in nano Ampere.
Voltage at the charger needle in Volts.
Pos. Charger Current
and Voltage
Ion current of the positive charger in nano Ampere.
Voltage at the charger needle in Volts.
External Inputs
Analog 1
Voltage of the external analog input Analog In CH1 in
Volts.
Analog 2
Voltage of the external analog input Analog In CH2 in
Volts.
Trigger channel 1
Status of the external Trigger IN (On or OFF)
External Outputs
Analog Out
Activates the Analog Output Settings Dialog. The
Analog Out button is only displayed if the Model 3090
EEPS spectrometer has the analog out option.
Note:
The Instrument Status Dialog is inactive while the
Analog Out dialog is displayed.
Controls
Neg. Charger
Use this push button to turn on/off the negative charger.
Pos. Charger
Use this push button to turn on/off the positive charger.
Column Heater
Use this push button to turn on/off the column heater.
Turn the heater on when the instrument is in a cool (i.e.,
air conditioned) environment and sampling from a warm,
humid environment to prevent condensation that could
cause the electrometers to become noisy. Normally, the
heater should be turned off unless needed.
Flow Control
Use this push button to turn on/off the flow control.
Analyzer Voltages
Use this push button to turn on/off the analyzer
voltages.
Содержание SIZER EEPS 3090
Страница 2: ......
Страница 18: ...xvi Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 20: ...xviii Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 36: ...2 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 144: ...B 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 149: ...Serial Commands D 3 3 Enter the communications COM port 4 Enter the port settings described below and click OK...
Страница 184: ......