Analog Output
E-3
Where:
dN
= Particle number concentration
e
= Particle Density
K
= Weighting Constant
max
Dp,
min
Dp,
= summation of channels or partial channels within the overall size
limits specified
pc
D
= Particle diameter size channel limits
pw
D
= Particle diameter range (that can be less than one channel width)
lower
upper
pc
D
D
log
D
log
d
where D
upper
and D
lower
are the upper and lower channel size boundaries,
for the model 3090 dlogD
pc
is always 1/16
lower
upper
pw
D
D
log
D
log
d
where D
upper
and D
lower
are the upper and lower size boundaries (may be
less than one channel width)
n
p
D
= Particle diameter (midpoint of size range
– channel or partial
channel) to the n power
Normalizing Multiplier:
min
p
max
p
O
D
/
D
log
1
N
if Normalized mode
is selected
1
O
N
if Normalized mode is not selected
Output Voltage:
min
,
fs
max
,
fs
min
,
fs
s
fs
O
C
C
C
C
V
V
for linear mode
min
,
fs
max
,
fs
min
,
fs
s
fs
O
C
log
C
log
C
log
C
log
V
V
for log mode
min
,
fs
C
= Concentration minimum value (cannot be zero if log mode is used)
max
,
fs
C
= Concentration maximum value
fs
V
= Voltage full scale (5V or 10V)
Содержание SIZER EEPS 3090
Страница 2: ......
Страница 18: ...xvi Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 20: ...xviii Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 36: ...2 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 144: ...B 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 149: ...Serial Commands D 3 3 Enter the communications COM port 4 Enter the port settings described below and click OK...
Страница 184: ......