6-32
Model 3090/3090AK Engine Exhaust Particle Sizer
TM
Spectrometer
Table 6-3
Fatal Error Flags
Error
Code
Error Message
Suggested Action
16
Column Voltage Section 1 Too Low
Contact TSI
17
Column Voltage Section 1 Too High
18
Column Voltage Section 2 Too Low
19
Column Voltage Section 2 Too High
20
Column Voltage Section 3 Too Low
21
Column Voltage Section 3 Too High
32
Absolute Pressure Too Low
33
Absolute Pressure Too High
37
Instrument Matrix Readback Error
38
ELM Covariance Readback Error
39
ELM Offset Readback Error
40
ELM Gain Readback Error
41
Missed DSP Interrupt
42
DSP Command Code Timeout Error
43
DSP Reset Timeout Error
44
COM1 Xmit Buffer Overflow Error
45
Flowmeter EEPROM Error
Check all connectors on main pc-board.
46
Serial Communication Error (Set by
Software)
If using a long cable, test with a standard length RS-
232 cable. If used in an electrically noisy environment,
remove instrument to a less noisy environment and try
again.
48
DSP Timing Error
Contact TSI
49
DSP Command Code Error
51
Electrometer Gain Error
Contact TSI
52
Electrometer Chan Above Noise Limit
During instrument zeroing, the offset current and/or
RMS noise for one or more channels was very high.
Retake the zero with the instrument disconnected from
any other equipment, in a vibration-free area, or with
the filter provided in the accessory kit (1031252) on the
inlet. Clean the column or contact TSI.
53
DSP Instrument Matrix Error
Contact TSI
55
DSP Data Not Valid
Contact TSI
56
DSP Checksum Error 1
57
DSP Checksum Error 2
58
Electrometer PCB 1 Error
There is a 60-conductor ribbon cable between the column
and the large pc-board on the door of the cabinet. Make
sure all six of the small connectors on the column and the
large connector on the pc-board are plugged in and seated
firmly. Power off the instrument and power it back on. If the
error remains, contact TSI.
59
Electrometer PCB 2 Error
60
Electrometer PCB 3 Error
61
Electrometer PCB 4 Error
62
Electrometer PCB 5 Error
63
Electrometer PCB 6 Error
Содержание SIZER EEPS 3090
Страница 2: ......
Страница 18: ...xvi Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 20: ...xviii Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 36: ...2 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 144: ...B 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 149: ...Serial Commands D 3 3 Enter the communications COM port 4 Enter the port settings described below and click OK...
Страница 184: ......