Quick Start
3-7
Note that this checkbox always defaults to “off”. If using it for consecutive
runs, it must be checked each time.
Hide Invalid Values
Check this box to have the software hide any invalid data values in its
graphs and tables. An invalid data value is one that is received by the
instrument under a serious or fatal error condition. See Table 6-3 in
Chapter 6
for a detailed list of fatal errors.
Dilution Factor
To correct for a dilution factor to match measurement conditions, you can
add a dilution factor here and turn it on/off with the checkbox. For example,
if you used a 100:1 diluter before sampling into the instrument, the
concentration can be corrected back to the original concentration using a
dilution factor of 100. This will multiply the concentrations measured by the
instrument by a factor of 100 when shown on graphs and tables. The
original data is not changed and the dilution factor can always be turned
back off.
Auto-Export
Select the Export Data at End of Run checkbox to automatically export data
when the run is complete. Refer to
Chapter 5
for a complete description of
how to export data.
Instrument Matrix
The software displays the currently configured matrix on the instrument in
the combo box. To select an alternate matrix, use the drop-down arrow.
DEFAULT refers to the factory default matrix; “SOOT” and “COMPACT” are
other pre-defined matrices which are installed with the software. When you
select a matrix other than the one currently configured on the instrument,
that matrix is uploaded to the instrument by selecting the
Apply
or
OK
button (to close the dialog).
Comment
The Comment box allows you to store comments about the sample run.
The maximum number of characters is 255.
This information is stored with the sample data in the file and may be
displayed when the Properties dialog box is opened for this file. It is also
included in the exported data.
Содержание SIZER EEPS 3090
Страница 2: ......
Страница 18: ...xvi Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 20: ...xviii Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 36: ...2 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 144: ...B 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 149: ...Serial Commands D 3 3 Enter the communications COM port 4 Enter the port settings described below and click OK...
Страница 184: ......