F-8
Model 3090/3090AK Engine Exhaust Particle Sizer
TM
Spectrometer
Read Total Mass and Particle Concentration
Command
Response
Description
ATPM K0
ATPM E PM
Total particle mass for all particle size bins
measured over 1 second.
PM =
Particle Mass
Range:
0 to 9.9999E9 [µg/m³]
Format:
Float
Example: PM = 1.2345E6
ATPC K0
ATPC E PC
Total particle concentration for all particle size
bins measured over 1 second.
PC =
Particle Concentration
Range:
0 to 9.9999E9 [#/cm³]
Format:
Float
Example: PC = 1.2345E5
Read Particle Density Vector
Command
Response
Description
APDV K0
APDV E B1 B2 B3…B32
Read particle size density from
instrument
BX =
Particle density for bin X
Range: 0.00 to 10.00 [g/cm³]
Format: Float
Read Electrometer
Command
Response
Description
AELO K0
AELO E C1 C2 C3…C22
Read Electrometer Offset
C =
Electrometer offset current
for channel X
(C22: Bottom Electrometer)
Range: -9.9999E9 to 9.9999E9 [fA]
Format: Float
AELN K0
AELN E C1 C2 C3…C22
Read Electrometer RMS Noise
C =
Electrometer noise level for
channel X
(C22: Bottom Electrometer)
Range: -9.9999E9 to 9.9999E9
[fA
RMS
]
Format: Float
Read Serial Number & Firmware Version
Command
Response
Description
ARSN K0
ARSN E XXXXXXXX
Read serial number
XXXXXXXX [Serial Number]
Format: Text
ARFV K0
ARFV E XXXX YYYY
Read firmware version
XXXX [Main Firmware]
YYYY
[DSP Firmware]
Format: Text
Содержание SIZER EEPS 3090
Страница 2: ......
Страница 18: ...xvi Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 20: ...xviii Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 36: ...2 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 144: ...B 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 149: ...Serial Commands D 3 3 Enter the communications COM port 4 Enter the port settings described below and click OK...
Страница 184: ......