6-10
Model 3090/3090AK Engine Exhaust Particle Sizer
TM
Spectrometer
Figure 6-9
Remove the Screws of the Positive Charger Assembly
C l e a n i n g t h e C h a r g e r C o l u m n
The charger column should be cleaned according to Table 6-1 or when an
error code indicates there is a problem. This section provides two
procedures for cleaning the charger column: a basic procedure and a
detailed procedure. Perform the basic procedure first. If that does not
resolve the problem, perform the detailed procedure.
Note:
Make certain the cyclone is always mounted on the aerosol inlet. The
cyclone removes lint and large particles from the air stream and will
help keep the instrument clean.
If you have questions regarding the need to clean the charger column, refer
to “
Technical Contacts
” later in this chapter.
W A R N I N G
High voltage is accessible in several locations within this instrument.
Make sure you unplug the power source to the instrument before
performing maintenance procedures.
Содержание SIZER EEPS 3090
Страница 2: ......
Страница 18: ...xvi Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 20: ...xviii Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 36: ...2 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 144: ...B 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 149: ...Serial Commands D 3 3 Enter the communications COM port 4 Enter the port settings described below and click OK...
Страница 184: ......