4-2
Model 3090/3090AK Engine Exhaust Particle Sizer
TM
Spectrometer
Status Bar
The Status Bar indicates instrument status. Possible messages include:
Warming Up, Normal, etc. Refer to
Chapter 6
for a complete list of
messages and actions you should take.
Menu Options
There are three menu options.
Menu Option
Description
Cursor
View individual particle channels.
Menu
View user settings and diagnostics.
Pause/Resume
Start and stop data display.
Cursor
Select
CURSOR
to view information about a particular particle size value in
the graph. To view a specific value, use the control knob to highlight
CURSOR then press the knob. A bar will appear on the graph and data for
that size channel will be displayed in the concentration display region (see
below). Rotate the control knob to view data to the left or right. To stop
viewing data with the cursor, press the control knob again and turn the
cursor to select another option from the display.
Menu
Select
MENU
to view user setting and diagnostics information. To view
these items, highlight MENU with the control knob and press the knob. A
menu appears for User Settings and Diagnostics. Use the control knob to
highlight the item you want and press the control knob.
User Settings
The user settings (Table 4-1) let you change display settings and turn on/off
the pumps or chargers and zero the electrometer. The pumps can be
turned off to reduce noise and save power when the instrument is not being
used, but still powered on.
Table 4-1
User Settings
Setting
Options
Description
Y-Axis Setting
Autoscale;
1E4 to 1E9
Use this option to select the maximum
displayed scale for the Y-Axis.
(1E4 can only be selected in Linear mode)
Y-Axis Scale
Linear/Log 10
Use this option to change the Y-Axis scale
between Linear and Log 10 mode.
Flows
Off/On
Use this option to turn all flows on or off.
When Off is selected, the status bar
displays “Flows and Chargers Turned
OFF
.” There will be no draw at the sample
inlet. The chargers are turned off to protect
the electrodes from contamination.
Содержание SIZER EEPS 3090
Страница 2: ......
Страница 18: ...xvi Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 20: ...xviii Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 36: ...2 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 144: ...B 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 149: ...Serial Commands D 3 3 Enter the communications COM port 4 Enter the port settings described below and click OK...
Страница 184: ......