![TSI Instruments SIZER EEPS 3090 Скачать руководство пользователя страница 153](http://html1.mh-extra.com/html/tsi-instruments/sizer-eeps-3090/sizer-eeps-3090_operation-and-service-manual_1178156153.webp)
Serial Commands
D-7
D e t a i l e d C o m m a n d D e s c r i p t i o n
READ Commands
RDR
Read Data Record
RDR,0
Unpolled mode OFF (stop streaming binary data)
RDR,1
Unpolled mode ON (start streaming binary data)
RDR,2
Unpolled mode ON (start streaming binary data
when instrument receives signal on trigger input)
RDR,3
Unpolled mode ON (start streaming binary data)
RDR,4
Unpolled mode ON (start streaming text data)
See TSI for more detailed information on data structure.
RSR
Read Status Record
RSR
Read Status Record
RSR,#
Read Status Record (without comments)
Error Code
x2,x3; U_SHORT; [-]
Status Code
x1; U_SHORT; [-]
Sheath Flow
x4; FLOAT; [L/min]
Sample Flow
x5; FLOAT; [L/min]
Charger Flow
x6; FLOAT; [L/min]
Extraction Flow
x7; FLOAT; [L/min]
Absolute Pressure
x8; FLOAT; [mBAr]
Analyzer Voltage Top
x9; FLOAT; [V]
Analyzer Voltage Middle
x10; FLOAT; [V]
Analyzer Voltage Bottom
x11; FLOAT; [V]
Sheath Flow Temp.
x12; FLOAT; [C]
Neg. Charger Current
x13; FLOAT; [nA]
Pos. Charger Current
x14; FLOAT; [nA]
Neg. Charger Voltage
x15; FLOAT; [V]
Pos. Charger Voltage
x16; FLOAT; [V]
Analog Input 1
x17; U_SHORT; [V]
Analog Input 2
x18; U_SHORT; [V]
RPID
Read PID Parameter
RPID
Read PID (n text)
RFV
Read Firmware Version
RFV
Firmware Version (text)
Содержание SIZER EEPS 3090
Страница 2: ......
Страница 18: ...xvi Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 20: ...xviii Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 36: ...2 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 144: ...B 8 Model 3090 3090AK Engine Exhaust Particle Sizer TM Spectrometer This page intentionally left blank...
Страница 149: ...Serial Commands D 3 3 Enter the communications COM port 4 Enter the port settings described below and click OK...
Страница 184: ......