7
•
Maintenance and Service
Page 126
Vanquish Diode Array Detectors (VH-D10) Operating Manual
7.7
Flow Cell
This section describes exchange, storage and shipment of LightPipe flow
cells.
To store or ship the flow cell, follow these steps:
1. Flush the flow cell (see
2. Remove the flow cell from the detector and place it in its packaging
(see
To exchange the flow cell, follow these steps:
1. Remove the flow cell from the detector (see
section 7.7.2,
page 128
2. Install the new flow cell to the detector (see
section 7.7.3,
page 130
7.7.1
Flushing the Flow Cell
When
Before the flow cell is stored in its packaging or in the detector (if
flushing the flow cell with highly pure isopropanol through the system is
not possible) or if it is shipped.
Parts required
•
Manual inject and flush kit for LightPipe flow cells, including a
syringe and a Viper adapter
•
Highly pure solvent that is miscible with isopropanol, such as LC/MS-
grade acetonitrile or methanol
•
Highly pure isopropanol, such as LC/MS-grade isopropanol
•
Plugs for the flow cell inlet and outlet (supplied with the flow cell)
Содержание Vanquish VH-D10
Страница 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Страница 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...
Страница 196: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...