5
•
Installation
Vanquish Diode Array Detectors (VH-D10) Operating Manual
Page 75
7. Read the system pressure in Chromeleon and write down the value
for P1.
8. Stop the pump flow.
9. Disconnect the waste line from the Viper union.
WARNING—Escape of Hazardous Substances from Flow Connections
Flow and capillary connections can be filled with substances that can
pose health risks. Solvent can spray when capillaries burst, slip out of
their fittings, or are not properly tightened or when capillary connections
are otherwise open.
•
Wear appropriate protective equipment and follow good laboratory
practice.
•
Before starting maintenance or repair procedures, flush out harmful
substances with an appropriate solvent.
10.
Only if a second detector is used
: Connect the waste line to the
outlet port of the second detector.
11. Connect the inlet port of the mass spectrometer/second detector to
the free end of the Viper union with the capillary which will be used
to connect the second detector or the mass spectrometer in the
application setup.
12. Start the pump flow at the maximum flow rate of your application.
13. Check all flow connections for leakages:
♦
If a leakage has occurred
: See
section 8.4 Resolving Leakage,
page 172
♦
If no leakage was found and the pressure value has stabilized
(approximately after 30 seconds): Proceed with the next step.
14. Read the system pressure in Chromeleon and write down the value
for P2.
15. Stop the pump flow.
Содержание Vanquish VH-D10
Страница 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Страница 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...
Страница 196: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...