5
•
Installation
Page 50
Vanquish Diode Array Detectors (VH-D10) Operating Manual
Power Cord
The power cords are designed to match the wall socket requirements of
the country in which they are used. The end of the power cords that
plugs into the power socket on the device is identical for all power cords.
The end of the power cords that plugs into the wall socket is different.
WARNING—Electric Shock or Damage to the Device
•
Never use a power cord other than the power cords provided by
Thermo Fisher Scientific for the device.
•
Only use a power cord that is designed for the country in which you
use the device.
•
Do not use extensions cords.
•
Never plug the power cord to a power socket that is shared with
other equipment (for example, multiple sockets).
•
Operate the device only from a power outlet that has a protective
ground connection.
•
In case of emergency, it must be possible to reach the power cord
easily at any time to disconnect the device from the power line.
WARNING—Electric Shock or Damage to a Product
Misuse of the power cords could cause personal injury or damage the
instrument. Use the power cords provided by Thermo Fisher Scientific
only for the purpose for which they are intended. Do not use them for
any other purpose, for example, for connecting other instruments.
Condensation
NOTICE
Condensation in the device can damage the electronics and
optics.
Therefore, when using, shipping, or storing the device, avoid or minimize
conditions that can lead to a build-up of condensation in the device. For
example, avoid significant or fast changes in environmental conditions.
If you suspect that condensation is present, allow the device to warm up
to room temperature. This may take several hours. Wait until the
condensation is gone completely before connecting the detector to the
power line.
Содержание Vanquish VH-D10
Страница 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Страница 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...
Страница 196: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...