6
•
Operation
Vanquish Diode Array Detectors (VH-D10) Operating Manual
Page 95
Parameters
Description
Wavelength
Set a wavelength for each
UV_VIS_x
signal channel
individually, in the range of 190 nm to 680 nm for up to
10 signal channels.
The wavelength defines the value at which the detector
measures the absorbance of the analyte(s) in the sample.
Observe the following:
•
Set the
Wavelength
to the wavelength with the
absorbance maxima for the analytes of interest.
•
For best selectivity, select a specific wavelength for every
substance to be analyzed with a low optical bandwidth,
for example, by using a separate signal channel for each
analyte.
•
To obtain optimum linearity, set the
Wavelength
to a
peak or a valley in the absorption spectrum. A valley can
provide best linearity for high concentrations.
Data collection rate
The data collection rate is the number of data points per
second (Hz) that Chromeleon collects from the detector
and stores as raw data. Select a data collection rate up to
200 Hz.
Set the data collection rate, at which data is to be
collected from the detector. The data collection rate
applies for all signal channels and the 3D field.
As a standard, when setting the
Data Collection Rate
parameter, the response time and peak width are defined.
For guidelines on selecting the data collection rate, see
section 6.7.3 Selecting the Data Collection Rate, page 102
Bandwidth
The bandwidth specifies the optical bandwidth (i.e. the
ability of the detector to distinguish between single
wavelengths), at which the signal channel (
UV_VIS
) is
recorded. In general, this corresponds to the optical
resolution of a detector.
The standard
Bandwidth
is set to 4 nm. If required, you
can set a different value for the bandwidth.
For details on the bandwidth and setting it to a different
value, see
Содержание Vanquish VH-D10
Страница 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Страница 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...
Страница 196: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...