7
•
Maintenance and Service
Vanquish Diode Array Detectors (VH-D10) Operating Manual
Page 139
Forward flushing
1. Switch the capillary connections on the flow cell: Connect the Viper
connector from the back-flush capillary to the flow cell inlet and the
back-flush waste line to the flow cell outlet.
Figure 35: Setup for forward flushing
No.
Description
1
To the outlet of the pump
2
Back-flush waste line
3
Waste container
2. Turn on the pump flow at the flow rate as determined in
section
Determining the required pump flow
3. Flush the flow cell for 2 minutes with the back-flushing agent.
4. Monitor the operating pressure:
♦
If the pressure remains high at a value similar to the value that
you wrote down, the clogging was not removed. Thermo Fisher
Scientific recommends performing another backward and
forward flushing cycle:
a) Stop the pump flow.
b) Follow the steps starting with
Back-flushing on page 138
If repeated flushing did not remove the clogging, proceed
with
Ending the flushing procedure below
.
♦
If the pressure is decreasing to a normal state and only a little
liquid flows through the overpressure line, the clogging is
removed from the flow cell. As the blockage is flushed out, the
pressure at the pump is usually decreasing. More of the liquid
will flow through the flow cell and less fluid is leaving the
overpressure line.
Proceed with the next steps.
Содержание Vanquish VH-D10
Страница 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Страница 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...
Страница 196: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...