9
•
Specifications
Vanquish Diode Array Detectors (VH-D10) Operating Manual
Page 177
9.1.2
LightPipe Flow Cells
The flow cell performances are specified as follows:
Type
10 mm Fused-Silica LightPipe Flow
Cell Specification
60 mm Fused-Silica LightPipe Flow
Cell Specification
Design
Light pipe
Light pipe
Path length
10 mm
60 mm
Illuminated volume
2 µL
13 µL
Dispersion volume
0.8 µL
4.0 µL
Pressure limit
6 MPa
6 MPa
Max. eluate
temperature limit
50 °C
50 °C
Biocompatibility
yes
yes
Materials in the
analytical flow path
Fused silica, PEEK, perfluoro-
elastomer, titanium
Fused silica, PEEK, perfluoro-
elastomer, titanium