5.6
Setting Up the Flow Connections
5.6.1
General Information and Guidelines
When setting up flow connections, follow these rules and
recommendations:
Flow connections can be filled with hazardous substances. Observe the
warning messages and precautionary statements presented in
section 2.3 Safety Precautions
page 20
NOTICE
Particulate matter from other system modules and components
can deposit in the flow cell and clog it. Before you connect the flow cell
to the flow path, make sure that you thoroughly flush the modules in the
system flow path upstream of the detector to waste.
When you install devices or components to the system, always flush
them to waste before connecting them in the system flow path. To flush
the Vanquish modules, follow the instructions in the
Vanquish System
Operating Manual
.
•
Dirty components can contaminate the chromatography system.
Contamination leads to poor performance of the modules and entire
system or can even cause damage to the modules and system.
Therefore:
♦
Always wear appropriate gloves.
♦
Place the components only on a clean, lint-free surface.
♦
Keep your tools clean.
♦
Use only lint-free cloth for cleaning.
•
For installation instructions and guidelines and for handling
recommendations, see
section 5.6.3 Connecting Fittings, Capillaries,
and Tubing, page 64
To set up the flow connections and complete the installation of the
detector, follow these steps:
1. Set up the flow connections to the flow cell (see
section 5.6.4,
page 67
2. Connect the detector to the drain system (see
section 5.6.5,
page 72
Содержание Vanquish VH-D10
Страница 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Страница 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...
Страница 196: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...