7
•
Maintenance and Service
Page 130
Vanquish Diode Array Detectors (VH-D10) Operating Manual
b) Position the cover onto the flow cell opening. Make sure
that the cover is in the correct orientation.
c) Turn the rotating locks clockwise to a vertical position to
close the rotating locks and attach the cover to the flow cell
opening.
NOTICE
The flow cell opening on the detector is sensitive to dust
and debris. If no flow cell is installed to the detector, close the flow
cell opening with the cover for the flow cell opening.
7.7.3
Installing the Flow Cell
When
•
Exchanging a flow cell
•
Installing a flow cell after storage
Parts required
LightPipe flow cell
NOTICE—Sensitive LightPipe Flow Cells
LightPipe flow cells are highly sensitive to damage. Handle LightPipe flow
cells with care. Observe the following notes when installing the flow cell
to the detector:
•
Mechanical shocks, mechanical vibrations or intruding objects can
lead to leaks on the flow cell or even destroy it. Avoid exposure of
the flow cell to mechanical shocks or vibrations. Do not let it hit hard
surfaces. Do not intrude the flow cell enclosure with any objects.
•
The optical ports of the flow cell are sensitive to contamination and
scratches. Do not touch the optical ports of the flow cell or immerse
them. To avoid damage to the optical ports of the flow cell, be
careful when inserting the flow cell into the flow cell opening of the
detector.
•
On the rear side of the flow cell, the contact pads for the
identification chip are located. Never touch the contact pads. Avoid
damage to the electronics of the ID chip.
Содержание Vanquish VH-D10
Страница 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Страница 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...
Страница 196: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...