Controlled-Rate Freezer
Active Run Summary | 31
Active Run Summary
A Profile Run can be initiated from the
tab or from
the
button on the Home screen.
Initiate a Run
1. Load the product into the freezing chamber.
2. Select the
tab.
3. Select the desired profile and press the Run button you
will be navigated to the Run Information Screen.
4. Enter run information as needed and press the Start Run.
Figure 79. Enter Run Information Screen
5. Once the profile is started, the profile run screen will be
displayed as shown in
.
Figure 80. Profile Run Screen.
The mode indicator will show the status and current step of
the profile.(Running: Preset_1, Step 3)
Note:
When a profile is initiated and the user is logged out
they cannot perform other operations on the screen.
Note:
When a profile is running the left navigation pane is not
visible.
You may zoom in and out in the chart region.
You may also pan tab views on the right pressing the left or
right arrows on the Profile Screen section / or by swiping
left or right.
•
Profile:
This view lists the steps of the actively running
profile.
•
Legend:
This view associates the series name and color.
•
Information Fields:
This view gives you the information
about Date/Time, User, Data Tag, Profile Name, Run File
Name, Note and any custom fields which you have
entered before running a profile.
•
Alarm Details:
This view lists all alarms that have
occurred during the run, up to the current time.
When a profile is running and any alarm occurs, icons of that
specific alarm will be displayed on the screen below the graph
lines.
When you press the icon you can see which alarm has
occurred. Refer
section for more information
Manual Override
The user may override a running profile by manually injecting
LN
2
or activating the heater to alter the cooling or heating
effect.
1. During a running profile, the unit will enable the functions
listed below provided a user is logged in.
2. Select the desired function to override (Warm, Cool or
Cool+) by pressing the specific icons.
CAUTION:
The manual override function
deviates from the profile and should not be used
to achieve cabinet temperatures below -180°C.
Manual override events are not recorded in run
log, but are recorded in the event log.
Advance:
Advances to next step of the
profile.
Cancel Run:
Cancels the run. User
must confirm the selection.
Cool:
Energizes the cool solenoid.