Troubleshooting
Troubleshooting Guide
Thermo Fisher Scientific
TVA2020 Instruction Manual
6-5
Problem
Possible Reason
Solution/Action
Service
pump.
Sampling
non-volatile
compounds
Call Thermo Fisher Scientific
Service Department.
High background readings High ambient concentration
N/A.
Zero
drift/improper
calibration
Ensure proper zero/span
calibration.
Sample line contamination
Clean/replace sample line.
Clean/replace filter cups.
Detector
capsule
contamination
Clean/replace FID capsule.
Clean PID window.
Replace PID capsule.