Display Menus
Run Mode
3-4
TVA2020 Instruction Manual
Thermo Fisher Scientific
How the instrument functions in the RUN mode is governed by the
selection made in the SETUP/Log menu. The LOG menu selects how data
is stored in the instrument memory. There are four possible choices in the
LOG mode as defined in “
Log Methods
” on page
3-15
. In addition, if the
optional Bluetooth module is active, data automatically streams out
continuously from the instrument. Data will only stream out of the
instrument while in the RUN mode. The data will always stream at in the
same format independent of the log configuration. (See “
Bluetooth
Communications
” in the Options section for more information.)
The TVA2020 performs accurately only after it has been properly set up
(configured).
The RUN mode displays (as governed by the LOG Selection) are:
FID-NONE
PID/FID-NONE
FID: 1376ppm PID: 1042ppm
FID: 1376ppm
Exit=Stop Exit=Stop
FID-AUTO, VOC, or F.E. PID/FID-AUTO, VOC, or F.E.
FID: 1376ppm PID: 1042ppm
FID: 1376ppm
Sampling: 3sec Sampling: 3sec
Exit=Stop Exit=Stop
You should install the hydrogen tank and wait 2-3 minutes before entering
the RUN mode. Upon entering the RUN mode, the pump will
automatically turn ON and the FID will ignite.
Note
For best performance, the instrument must be ON and warmed up
for approximately 30 minutes with the FID flame ignited throughout the
warmup period. If the instrument is PID equipped and the PID is to be
used, the lamp must be ON throughout the warmup period.
▲
Run Mode
Содержание TVA2020
Страница 1: ...TVA2020 Instruction Manual Toxic Vapor Analyzer Part Number 111755 00 29Jul2014...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 16: ...Tables xii TVA2020 Instruction Manual Thermo Fisher Scientific...
Страница 76: ......
Страница 96: ......
Страница 118: ......
Страница 178: ...Text Data Formats Factory B 14 TVA2020 Instruction Manual Thermo Fisher Scientific SUM OF SQUARES 0 934123 END...
Страница 190: ...Response Factors Response Factors C 4 TVA2020 Instruction Manual Thermo Fisher Scientific...