Preventive Maintenance
Hydrogen Gas Tank
Thermo Fisher Scientific
TVA2020 Instruction Manual
5-5
clockwise) until the rubber tank boot is flush with the instrument sidepack
and a slight resistance is felt.
Do not overtighten
.
WARNING
A safe refill operation means there are no hydrogen leaks.
Before any valves are opened, use a wrench to firmly tighten connections to
the hydrogen supply tanks and the tank fill adapter. If escaping hydrogen is
heard during the filling operation, close all valves and correct the leak
before proceeding. Leak test with soapy water or equivalent.
▲
CAUTION
Do not fill hydrogen tank to a pressure greater than 15.2 MPa
(2200 psig).
▲
Note
Use prepurified ZERO grade hydrogen (certified total hydrocarbons
as methane <0.5 ppm recommended).
▲
Note
The hydrogen filling assembly contains a flow-limiting safety device.
Approximately two minutes are required to fill the tank.
▲
The major hazard associated with the handling of hydrogen is flammability.
The following specific rules apply when handling hydrogen:
●
Never use cylinders of hydrogen in areas where flames, excessive heat,
or sparks may occur.
●
Use only explosion-proof equipment and spark-proof tools in areas
where hydrogen is handled.
●
Ground all equipment and lines used with hydrogen.
●
Never use a flame to detect hydrogen leaks — use soapy water.
●
Do not store reserve stocks of hydrogen with cylinders containing
oxygen or other highly oxidizing or combustible materials.
●
Store hydrogen tanks in a well ventilated area.
●
Follow all regulatory safety and labeling precautions when shipping
hydrogen in the TVA2020.
Note
A bleeder assembly is provided in the tool kit to allow the hydrogen
tank to be emptied for common carrier shipment. To use the bleeder,
manually screw the bleeder onto the tank valve and allow the hydrogen to
vent. This process will take about two minutes to complete.
▲
Precautions in
Handling and
Storage
Содержание TVA2020
Страница 1: ...TVA2020 Instruction Manual Toxic Vapor Analyzer Part Number 111755 00 29Jul2014...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 16: ...Tables xii TVA2020 Instruction Manual Thermo Fisher Scientific...
Страница 76: ......
Страница 96: ......
Страница 118: ......
Страница 178: ...Text Data Formats Factory B 14 TVA2020 Instruction Manual Thermo Fisher Scientific SUM OF SQUARES 0 934123 END...
Страница 190: ...Response Factors Response Factors C 4 TVA2020 Instruction Manual Thermo Fisher Scientific...