Calibration
Calibration Configuration
Thermo Fisher Scientific
TVA2020 Instruction Manual
4-7
reading will be the value displayed and the value logged. Choosing this
selection from the CAL CONFIG MENU produces the following display:
FID
PID/FID
Backgnd correct: Backgnd correct:
ON PID
1=Both 2=None
1=ON 2=None 3=PID 4=FID
The second line of the display shows what option is currently selected. You
can change this option by choosing any one of the menu selections in dual
detector instruments. You may choose to apply correction to FID, PID or
both detectors.
From the CALIB MENU display, press 3=Backgnd.
FID
PID/FID
FID: 0.0ppm PID: 0.0ppm
FID: 0.0ppm
Background conc Background conc
1=Update 2=Zero 1=Update 2=Zero
Press 1=Update to read live values of ambient air.
FID
PID/FID
Please wait... Please wait...
Sampling backgnd Sampling backgnd
Enter=Accept Enter=Accept
FID
PID/FID
FID: 0.3ppm PID: 0.2ppm
FID: 0.3ppm
Sampling backgnd Sampling backgnd
Enter=Accept Enter=Accept
Press ENTER.
Defining the Background
Correction