Display Menus
Log Methods
3-20
TVA2020 Instruction Manual
Thermo Fisher Scientific
1.
From the LOG MODE menu, press 3=VOC/F.E. The previous log
selection is displayed on the second line of the screen.
LOG MODE
Auto,rate= 10sec
1=VOC 2=FE
Press 2=FE to display this message.
Enter sampling:
FE,time=00s,Max
Up/Dn=Next type
Enter=Accept
2.
Press the Up/Dn arrow keys to choose the type of sampling to be used
(Max, Avg, Smp).
a.
Max = the maximum reading obtained during the log interval.
b.
Avg = the average reading obtained during the log interval.
c.
Smp = the last reading obtained during the log interval.
3.
Type the appropriate time interval desired for F.E. sampling.
4.
Press Enter, and, after a brief message of acceptance, the F.E. method
of logging is now ready to be used in the RUN mode.
Note
The allowable Sample Time range is 2-30 seconds. If a Sample Time
of less than 2 seconds or greater than 30 seconds is selected, a warning
message will appear and you will be required to re-enter the Sample
Time.
▲
Note
If dual option, the user must use the enhanced probe to change
detector type.
▲
Содержание TVA2020
Страница 1: ...TVA2020 Instruction Manual Toxic Vapor Analyzer Part Number 111755 00 29Jul2014...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 16: ...Tables xii TVA2020 Instruction Manual Thermo Fisher Scientific...
Страница 76: ......
Страница 96: ......
Страница 118: ......
Страница 178: ...Text Data Formats Factory B 14 TVA2020 Instruction Manual Thermo Fisher Scientific SUM OF SQUARES 0 934123 END...
Страница 190: ...Response Factors Response Factors C 4 TVA2020 Instruction Manual Thermo Fisher Scientific...