Calibration
Flow Calibration
Thermo Fisher Scientific
TVA2020 Instruction Manual
4-19
a.
Typical values are 65-75% power with 29K-31K counts
8.
Press
Enter=Accept
to save the Flow 2 values.
9.
Attach a second red flow restrictor, in line with the first, between the
probe inlet and flow meter.
10.
Press “
3
” for the third flow calibration point.
11.
Adjust the pump power with the up and down arrows for a reading of 1
±.01 LPM at the probe inlet.
a.
Typical values are 85-95% power with 31K-33K counts.
12.
Press
Enter=Accept
to save the Flow 3 values.
13.
Press
4= calc coef
to calculate the flow calibration values.
Note
The flow calibration values are stored in a file called
FACTORY.TXT. This file can be saved to a computer and reloaded to the
instrument if the flow calibration were to become corrupted. See the
section on flash file transfers.
▲
Содержание TVA2020
Страница 1: ...TVA2020 Instruction Manual Toxic Vapor Analyzer Part Number 111755 00 29Jul2014...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 16: ...Tables xii TVA2020 Instruction Manual Thermo Fisher Scientific...
Страница 76: ......
Страница 96: ......
Страница 118: ......
Страница 178: ...Text Data Formats Factory B 14 TVA2020 Instruction Manual Thermo Fisher Scientific SUM OF SQUARES 0 934123 END...
Страница 190: ...Response Factors Response Factors C 4 TVA2020 Instruction Manual Thermo Fisher Scientific...