Calibration
Calibration Scenarios
4-2
TVA2020 Instruction Manual
Thermo Fisher Scientific
If, instead, you want to measure in direct readings (response factor = 1) for
one specific compound with maximum accuracy over a wide range of
concentrations, perform a multipoint calibration with the specific
compound. Up to 9 span points (plus zero) can be entered for each
detector. The use of a response curve is thus unnecessary as the detector is
already reading the direct ppm for that specific compound.
CAUTION
If you use multipoint calibration or a gas other than methane or
isobutylene and then apply response factors/curves (that have been
generated with reference to a single point methane/isobutylene calibration),
the resulting measurements will probably be incorrect.
▲
To provide the specified accuracy, the instrument must be calibrated at the
beginning of each workday.
●
From the Main Menu, choose 2 = Setup >
1 = Calib
.
When you reach the CALIBRATION menu, you will see the following
selections:
CALIB MENU
1=Zero 2=Span
3=Backgnd 4=RF
5=Flow 6=Cfg
The steps involved in calibrating the TVA2020 are as follows:
1.
Configure the calibration variables (6=Cfg):
●
Number or span points (1-9)
●
Span concentration
●
Background correction (None, PID, FID, Both)
●
Response factor calculation mode (Multiplication factor or Curve)
●
Accept mode (Manual or Auto)
●
Save mode (Manual or Auto)
2.
Zero the instrument using either a zero gas or clean ambient air
(1=Zero).
Scenario 3
Содержание TVA2020
Страница 1: ...TVA2020 Instruction Manual Toxic Vapor Analyzer Part Number 111755 00 29Jul2014...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 16: ...Tables xii TVA2020 Instruction Manual Thermo Fisher Scientific...
Страница 76: ......
Страница 96: ......
Страница 118: ......
Страница 178: ...Text Data Formats Factory B 14 TVA2020 Instruction Manual Thermo Fisher Scientific SUM OF SQUARES 0 934123 END...
Страница 190: ...Response Factors Response Factors C 4 TVA2020 Instruction Manual Thermo Fisher Scientific...