Preventive Maintenance
Replacing Water Trap Probe Filter and O-Rings
Thermo Fisher Scientific
TVA2020 Instruction Manual
5-11
Use the following procedure to replace the water trap probe filter (
Figure
5–4
).
1.
Unscrew the bottom housing from the top housing.
2.
Remove filter.
3.
Place new filter into cavity of top housing.
4.
Filter can be inserted in either direction into top housing.
5.
Screw bottom housing into top housing.
Figure 5–4.
Water Trap Probe Assembly – Filter and O-Ring Replacing
Replacing Water
Trap Probe Filter
and O-Rings
O-ring
Nozzle Base
O-ring
Nozzle
Probe Nut
Water
Trap