Calibration
Defining the Response Factor
Thermo Fisher Scientific
TVA2020 Instruction Manual
4-17
When you press 2=Span, the display will show:
FID
PID/FID
FID: 44750 PID: 4000
FID: 44750
Span Pt1 (Up/Dn) Span Pt1 (Up/Dn)
Enter=New calib 1=Both 2Pid 3Fid
If more than one reference span gas concentration has been configured,
pressing the Up/Dn keys will allow you to scroll through all defined
reference gas concentration points.
2.
To perform the actual SPAN calibration (for example at 100 ppm
isobutylene):
●
FID-only: Press ENTER
●
PID/FID: Press 1, 2, or 3
FID
PID/FID
Apply span gas1 Apply span gas1
100.0ppm 100.0ppm
Enter=Start Enter=Start
3.
Apply the appropriate span gas to the probe at ambient pressure (using
a clean and labeled gas sampling bag) and then press ENTER.
FID
PID/FID
FID: 44750 PID: 4000
FID: 44750
Wait for span... Wait for span...
Exit=Cancel Exit=Cancel
The instrument analyzes the span sample.
When only one reference span gas concentration is used, the ACCEPTED
message appears for a short time and is then replaced by the normal
CALIBRATION menu.
Содержание TVA2020
Страница 1: ...TVA2020 Instruction Manual Toxic Vapor Analyzer Part Number 111755 00 29Jul2014...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 16: ...Tables xii TVA2020 Instruction Manual Thermo Fisher Scientific...
Страница 76: ......
Страница 96: ......
Страница 118: ......
Страница 178: ...Text Data Formats Factory B 14 TVA2020 Instruction Manual Thermo Fisher Scientific SUM OF SQUARES 0 934123 END...
Страница 190: ...Response Factors Response Factors C 4 TVA2020 Instruction Manual Thermo Fisher Scientific...